Customization: | Available |
---|---|
After-sales Service: | Factory After Serice by Real Time |
Warranty: | 1 Year |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
01 | 02 |
Stainless steel material Corrosion and oxidation resistance Use more durable materials |
Suppressing easily Multiple measuring ranges Customizable according to requirements |
1 | Output port 1 | Connect standard table or inspected table |
2 | Output port 2 | Connect standard table or inspected table |
3 | Output port 3 | Connect standard table or inspected table |
4 | Output port 4 | Connect standard table or inspected table |
5 | Pneumatic piston | Gas generating device |
6 | Pneumatic lever | Drive the piston rod to move and generate pressure through operation |
7 | Fine adjustment of air pressure | Increase or decrease pressure within a small range |
8 | Positive and negative pressure switching valve | Press in to measure positive pressure, pull out to measure negative pressure |
9 | Air pressure relief valve | Close during pressurization and release gas at the end of measurement |
10 | Pneumatic pump body | Combination installation of various components |
11 | Nameplate | Product Name Model Specification |
12 | Hydraulic pressure rod | Increase pressure through operation |
13 | Injection port | Add purified water or calibration oil |
14 | Hydraulic pressure relief valve | Close during pressurization and open during depressurization |
15 | Hydraulic vernier | Increase or decrease pressure within a small range |
Product name | Gas liquid dual-use pressure calibration table |
Model | ALKT785 |
Pressure range | Air pressure -0.05~60Bar/Hydraulic pressure 0~600Bar/0~1000Bar |
Measurement medium | Clean air/purified water orTransformer oil 25# (choose one) |
Way of pressure | Hand movement |
Ambient temperature | 0~50ºC |
Output port | Four |
Thread specification | Quick connect thread (M20 × 1.5) |
Cover form | Air pressure min 10Pa/hydraulic pressure min 0.01bar |
Product size | 500×410×195mm |
Product weight | Approximately 14kg |
Package size | 530×440×225mm(packaging material is cardboard box) |
Package weight | Approximately 16kg |
Product standard configuration | Certificate of conformity x 1 Instruction manual x 1 Warranty card x 1 Sealing ring x 4 4 plugs |
Working principle and usage |
Hydraulic pressure |
Unscrew the plug on the hydraulic output port and adjust the fine adjustment to the appropriate position (middle position). Close the pressure relief valve and gently operate the lever to allow the pressure guiding liquid to reach the two pressure output ports. Reinstall the standard gauge and the inspected gauge, operate the pressure rod, apply pressure to around 300Bar, use fine-tuning to adjust to the required pressure, and check point by point until the full range is reached. After inspection, open the pressure relief valve to reset the pressure to zero. After the pressure relief is completed, the inspected table and standard table can be removed. |
Air pressure |
Positive pressure: Unscrew the plug on the air pressure output port, install the standard gauge and the calibrated gauge, push the positive and negative pressure switching valve into the positive pressure position, close the pressure relief valve, rotate the micro adjustment counterclockwise to the farthest position, and use the pressure rod to apply the required pressure. If the required pressure is too high and the pressure of the inflation rod cannot be reached, you can use the micro adjustment to adjust it to your desired pressure. After calibrating and releasing the pressure, you can remove the calibrated gauge and the standard gauge. Negative pressure: Install the standard gauge and the calibrated gauge, pull out the positive and negative pressure switch to the negative pressure position, rotate the micro adjustment counterclockwise to the maximum position, close the pressure relief valve, and use the pressure rod to apply pressure to your desired pressure. When the pressure is close to the desired value, use the micro adjustment to adjust to your desired pressure. After calibrating the gauge and releasing the pressure. Remove the inspected table and standard table. |